Set up an intravenous infusion
Set up and run through an intravenous infusion; be aware of the different equipment and devices used.
Target · Perform under supervision0% ready
Do it on a patient with a supervisor observing you (GMC “direct supervision”).
Overview
Prime the line, exclude air, connect to a patent cannula, and set an accurate rate against the prescription.
Key numbers
- Drip rate (drops/min) = volume(mL) × drop factor ÷ time(min). Standard giving set = 20 drops/mL (blood set = 15)Manufacturer
- Routine maintenance fluids ≈ 25–30 mL/kg/day water, with ~1 mmol/kg/day Na⁺/K⁺/Cl⁻ and 50–100 g/day glucoseNICE CG174
Indications
- IV fluid resuscitation/maintenance
- IV medication administration
Contraindications & cautions
- Check the prescription (fluid, additives, rate), the bag (expiry, clarity, correct fluid) and the patient (fluid status, allergies)
Equipment
Prescribed fluid bagGiving set (± volumetric pump)Patent IV cannula, flushDrip stand
Step by step
Prepare
- 1Check the prescription and the fluid bag against it (right fluid, additives, expiry, no cloudiness)
- 2Close the roller clamp; spike the bag; hang it
Prime
- 1Squeeze the drip chamber to half-fill it
- 2Open the roller clamp and run fluid through the whole line to exclude all air, then close
- 3Confirm the cannula is patent (flushes without pain/swelling)
Run
- 1Connect to the cannula using ANTT
- 2Set the rate — via a pump, or by drops/min (drops/min = volume(mL) × drop factor ÷ time(min))
- 3Label the bag/line; monitor the site and the patient
Complications
- Air embolus, extravasation/infiltration, phlebitis, fluid overload, wrong rate
OSCE tips
- Exclude all air from the line — a safety marker.
- Right-check the prescription and the bag before connecting.
- Be able to calculate a manual drip rate.
Read
Sources & credits
Authored to UK finals standard and verified against UK guidance (Manufacturer · NICE CG174) and Geeky Medics OSCE guides. Southampton teaching materials © University of Southampton. Always follow current local policy.